SE-9639 Franck-Hertz Experiment
14
012-14264A
Experiment Procedure 2
Using a PASCO Interface and Data Acquisition Software
Items Needed
*See the PASCO web site at www.pasco.com for more information
Hardware Setup: Connect Cables and Cords
1. Connect all the cables and cords between the argon tube enclosure and the power supplies and current amplifier.
2. Connect one 8-pin DIN Extension Cable (UI-5218) from the INTERFACE port on the DC Current Amplifier to ANA-
LOG INPUT A on the Universal Interface (UI-5100).
Item* Quantity
850 Universal Interface (UI-5000) 1
PASCO Capstone Software (UI-5400) 1
850 Universal
Interface
Power Supply II
Current Amplifier
Power Supply I
Argon Tube
Enclosure
Analog
Input A
Analog
Input B
Interface Port
8-pin DIN Extension Cable
Note: Before connecting any cords or cables, be sure that all power switches on the
Interface, Power Supplies, and Current Amplifier are in the OFF position and all volt-
age controls are turned fully counterclockwise.
Comments to this Manuals