  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 17
Sodium Ion Selective Electrode 012–06615A
14
Symptom Possible Causes Next Step
Noisy or Unstable electrode exposed to interferences soak overnight in electrode storage solution
Readings (readings
continuously or rapidly defective electrode replace electrode
changing)
ISA not used use recommended ISA
stirrer not grounded ground stirrer
Drift (reading slowly samples and standards at different allow solutions to come to room temperature
changing in one direction) temperatures before measurement
electrode exposed to interferences soak overnight in electrode storage solution
incorrect reference filling solution use recommended filling solution
pH too acidic adjust pH with ISA
ISA not used use recommended ISA
Low Slope or No Slope standards contaminated or prepare fresh standards
incorrectly made
standard used as ISA use ISA
electrode exposed to interferences soak overnight in electrode storage solution
pH too acidic adjust pH with ISA
air bubble on membrane remove bubble by re-dipping probe